Model Number 7601-90003
Medical Device Problem Code Failure to Eject (4010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Lot unknown (2): devices not yet returned.Lot kfjx (2): manufacturing date 06 january 2020, devices returned 14 april 2022 the returned devices were was inspected visually and functionally.Visual inspection revealed wear consistent with regular use of the device but did not reveal any significant deformation or damage.Functional inspection during bench testing could not replicate the described failure mode.The inserter and a model screw were assembled; the hexalobe fit into the screw without difficulty.The screw was inserted into a model bone.After the screw was fully inserted into the model bone, the inserter was disengaged from the screw without difficulty.A conclusive root cause could not be determined with the available information.Factors which may have contributed to the event include the use of excessive force, use of the device at a difficult angle, hard bone quality of the patient, and general device wear.
 
Event or Problem Description
This record is a consolidation of q2 2022 records summarized as a part of the fda voluntary malfunction summary reporting program.This report captures four instances of intraoperative yukon screw inserter coupling issues.In all instances, the procedure was completed successfully with no surgical delay and no adverse consequence to the patient.
 
Additional Manufacturer Narrative
Lot unknown (2): devices not returned.Lot kfjx (2): manufacturing date 06 january 2020, devices returned 14 april 2022.
 
Event or Problem Description
This record is a consolidation of q2 2022 records summarized as a part of the fda voluntary malfunction summary reporting program.This report captures four instances of intraoperative yukon screw inserter coupling issues.In all instances, the procedure was completed successfully with no surgical delay and no adverse consequence to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW INSERTER
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key15135848
Report Number3004774118-2022-00302
Device Sequence Number5196479
Product Code LXH
UDI-Device Identifier10888857352605
UDI-Public10888857352605
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
Number of Events Summarized4
Summary Report (Y/N)Y
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7601-90003
Device Catalogue Number7601-90003
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/01/2022
Supplement Date Received by Manufacturer10/01/2022
Initial Report FDA Received Date07/29/2022
Supplement Report FDA Received Date10/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-