• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA; CATHETER PLACEMENT KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA; CATHETER PLACEMENT KIT Back to Search Results
Model Number N/A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of regn3972 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "repeated mechanical occlusions - picc replaced with sl picc, kink at 13 cm." additional information received 07/20/2022: it was reported "this picc was placed on (b)(6) - kink found on day ((b)(6)).The line was withdrawn 2 cm with resolution of kink and patency restored.(b)(6) picc occluded again and could not be restored, line was removed with kink showing at 13 cm.New picc placed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC PROVENA
Common Device Name
CATHETER PLACEMENT KIT
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15135886
Report Number3006260740-2022-02959
Device Sequence Number20135643
Product Code LJS
UDI-Device Identifier00801741213311
UDI-Public(01)00801741213311
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K053501
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model NumberN/A
Device Catalogue NumberCK000993
Device Lot NumberREGN3972
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/07/2022
Initial Report FDA Received Date07/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-