• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. HEALICOIL RSB SA 4.75MM W/1 UT & 1 UB BL; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. HEALICOIL RSB SA 4.75MM W/1 UT & 1 UB BL; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 72203697
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Rupture (2208); Device Embedded In Tissue or Plaque (3165)
Date of Event 07/07/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopy procedure, when the 4.75mm healicoil regenesorb suture anchor was inserted into the bone hole, the suture and tape came off the anchor.The anchor could not be removed from the body.The procedure was successfully completed placing a back-up device into an additional drilled bone hole.There was a non-significant surgical delay of 10min, and no further complications were reported.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.Insufficient product information was provided, thus a manufacturing record review could not be performed.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.This case reports, the suture and tape came off the anchor after insertion in the bone and reportedly, ¿the anchor could not be removed from the body.¿ no clinically relevant supporting documentation was provided for review.Based on the limited information provided the root cause of the breakage could not be determined.It is unclear if the ifu [10601068 rev j] were adhered to; however, use of excessive force during insertion can cause failure of the suture anchor or insertion device.Bone quality must be adequate to allow proper placement of the suture anchor.If more torque is required to insert the anchor, stop, and ensure that the hole size and depth are correct for the bone conditions encountered.If a portion of the anchor remains retained in the patient, it is made of a bio-composite material which is intended for implantation to allow for bone ingrowth.Since the anchor was retained, it is unknown if the anchor will migrate and there is potential risk for tissue inflammation and/or pain.No further clinical/medical assessment is warranted at this time.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALICOIL RSB SA 4.75MM W/1 UT & 1 UB BL
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15136065
Report Number1219602-2022-01077
Device Sequence Number14769320
Product Code MAI
UDI-Device Identifier00885554031393
UDI-Public885554031393
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K123393
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72203697
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2022
Supplement Date Received by Manufacturer10/10/2022
Initial Report FDA Received Date07/29/2022
Supplement Report FDA Received Date10/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-