BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24674
Medical Device Problem Codes Inflation Problem (1310); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that balloon rupture occurred.The severely stenosed target lesion was located in central vein.A 12.0 x 40, 135cm mustang balloon catheter was advanced for dilation.During initial inflation, at less working pressure for few seconds, the balloon failed to fully inflate, and burst during the process.The balloon did not detach inside the patient after it ruptured.The device was removed and the procedure was completed with another of same device.There were no patient complications reported.The patient was stable following the procedure.
 
Additional Manufacturer Narrative
E1: initial reporter facility name: (b)(6) hospital.Device evaluated by mfr.: mustang 12.0 x 40, 135cm was received for analysis.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.Blood was identified within the balloon which is evidence of a device leak.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 4mm distal of the proximal markerband.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.A visual examination observed no issues or damage to the tip of the device.A visual and tactile examination found no kinks or damage the shaft of the device.A visual examination observed no issues with the markerbands of the device.Both markerbands were undamaged and present on the shaft of the device.This concludes the product analysis.
 
Event or Problem Description
It was reported that balloon rupture occurred.The severely stenosed target lesion was located in central vein.A 12.0 x 40, 135cm mustang balloon catheter was advanced for dilation.During initial inflation, at less working pressure for few seconds, the balloon failed to fully inflate, and burst during the process.The balloon did not detach inside the patient after it ruptured.The device was removed and the procedure was completed with another of same device.There were no patient complications reported.The patient was stable following the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15136083
Report Number2124215-2022-25617
Device Sequence Number14769439
Product Code FGE
UDI-Device Identifier08714729793953
UDI-Public08714729793953
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K141521
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/06/2024
Device Model Number24674
Device Catalogue Number24674
Device Lot Number0027271552
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/06/2022
Supplement Date Received by Manufacturer08/26/2022
Initial Report FDA Received Date07/29/2022
Supplement Report FDA Received Date08/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
Patient Weight80 KG
-
-