Lot hhnp (1): manufacturing date 30 october 2018, device returned 30 june 2022 visual inspection confirms that the tip of the threaded inner shaft has fractured.There is slight rotational deformation on the hexalobe of the tip of the driver.This deformation indicates excessive torque may have been applied to the driver.Complaint history for the reported lot number was reviewed and no adverse trends were identified.Device history records were reviewed and no relevant manufacturing issues were identified.Limited information regarding the event was made available to stryker.Multiple attempts were made to obtain additional information, but no response was received.A root cause cannot be established.If additional information is received or the device is returned for evaluation, the investigation will be reopened and updated.Lot pjfj (1): device returned (b)(6) 2022.Visual inspection confirms that the tip of the threaded inner shaft has fractured.There is slight rotational deformation on the hexalobe of the tip of the driver.This deformation indicates excessive torque may have been applied to the driver.Complaint history for the reported lot number was reviewed and no adverse trends were identified.A conclusive root cause can not be determined.Limited information regarding the event was made available to stryker.Multiple attempts were made to obtain additional information, but no response was received.A root cause cannot be established.If additional information is received or the device is returned for evaluation, the investigation will be reopened and updated.
|