Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code Hematoma (1884)
Date of Event 06/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that when the catheter was paused, under ultrasound, the doctor had difficulty introducing the catheter into the jugular despite correct dilation.He used another catheter from the same lot, the patient has a jugular hematoma.Both devices were discarded after event.Additional information received 7/11/22: it was reported by healthcare professional ¿the hematoma occurred when the kit was attempted despite placement under echo and optimal dilation with, i think, a kit that had to go partly subcutaneously.So the hematoma is consecutive to the installation of the kit and prevented the installation of another kit in the same place.It was reported this occurred with two devices.This report addresses the first device.
 
Additional Manufacturer Narrative
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of refx3191 showed one other similar product complaint(s) from this lot number.The complaints for this lot number have been reported from the same facility in france.Device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIAGARA DIALYSIS CATHETER KIT 13.5F X 15CM
Common Device Name
CATHETER, HEMODIALYSIS, NON-IMPLANTED
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15136150
Report Number3006260740-2022-02975
Device Sequence Number14751646
Product Code MPB
UDI-Device Identifier00801741045608
UDI-Public(01)00801741045608
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K090102
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number5594150
Device Lot NumberREFX3191
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/11/2022
Initial Report FDA Received Date07/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-