• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS BD SAFETYGLIDE¿ NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS BD SAFETYGLIDE¿ NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Catalog Number 305921
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: date of event is unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that needle pulled out of hub while using bd safetyglide¿ needle.No patient impact reported.The following information was provided by the initial reporter: upon engaging the safety needle guard to cover the used needle, the needle detached from the securement base.No injury to staff or patient.
 
Event or Problem Description
It was reported that needle pulled out of hub while using bd safetyglide¿ needle.No patient impact reported.The following information was provided by the initial reporter: upon engaging the safety needle guard to cover the used needle, the needle detached from the securement base.No injury to staff or patient.
 
Additional Manufacturer Narrative
H6: investigation summary: two photos were provided to our quality team for investigation.Through visual inspection, it was observed that cannula disconnected from the hub.Based on the investigation with the photo sample analysis the symptom reported is confirmed.Since no physical sample displaying the reported condition was received a potential root cause could not be defined.A device history record review was completed for provided lot number 1204567.A review showed no rejected inspections or quality issues during the production that could have contributed to the reported defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SAFETYGLIDE¿ NEEDLE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15136319
Report Number1213809-2022-00444
Device Sequence Number5110623
Product Code FMI
UDI-Device Identifier00382903059218
UDI-Public00382903059218
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K951254
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/30/2026
Device Catalogue Number305921
Device Lot Number1204567
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2022
Supplement Date Received by Manufacturer08/18/2022
Initial Report FDA Received Date07/29/2022
Supplement Report FDA Received Date09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-