| Model Number |
1012453-12 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Rupture (1546); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/28/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion in the left anterior descending (lad) artery.After implanting a non-abbott stent, a 4x12mm nc trek balloon dilatation catheter (bdc) was used for post-dilatation.While attempting to remove the bdc and pull it back into the guide catheter (gc), the balloon separated into the aorta.A snare was used to remove the balloon without issue.There was no adverse patient sequela and no reported clinically significant delay in the procedure or therapy.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual and functional inspections were performed on the returned device.The reported balloon rupture was not confirmed; however, the reported separation was confirmed.The reported difficulty removing the device could not be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported balloon rupture and difficulty removing the device; however, the reported separation, unexpected medical intervention and removal of foreign body appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
Subsequent to the initial filed emdr report, the following information was received: the balloon ruptured at unknown atmospheres.When the device was pulled back into a guide liner, the balloon separated in the aorta.No additional information was provided.
|
| |
|