• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1804300-08
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was discarded and not returned.The lot history record (lhr) for this product could not be reviewed and a similar complaint query of the complaint handling system could not be performed because the product was not returned for evaluation and the lot number was not reported.The investigation determined the reported dislodged stent and subsequent treatment appear to be related to the operational context of the procedure.Additionally, the reported removal of foreign body was due to a dislodged stent within the anatomy.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat a heavily calcified lesion in the proximal left anterior descending (plad) artery.A 3x8mm xience skypoint drug eluting stent (des) was advanced through a xience stent that was implanted 2 days earlier.When pulling back to position the stent at the target lesion, the 3x8mm xience skypoint stent interacted with the previously implanted stent and dislodged from the balloon, distal to the implanted stent.A snare was used, and the dislodged stent was successfully removed.There was no adverse patient sequela and no reported clinically significant delay in the procedure or therapy.The lesion was ballooned with an unspecified balloon to successfully complete the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15136368
Report Number2024168-2022-08382
Device Sequence Number5120038
Product Code NIQ
UDI-Device Identifier08717648233210
UDI-Public08717648233210
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1804300-08
Device Catalogue Number1804300-08
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/07/2022
Initial Report FDA Received Date07/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-