• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. OXINIUM FEM HD 12/14 36 MM L/+8; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. OXINIUM FEM HD 12/14 36 MM L/+8; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71343608
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 01/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that the plaintiff suffered from a right hip dislocation and required a closed reduction under moderate sedation on (b)(6) 2022.X-ray demonstrated that the femur was displaced superior and lateral to the acetabular cup.The patient tolerated the procedure well.This patient has a history of a primary right hip r3-tha surgery on (b)(6) 2017 and a subsequent revision surgery on (b)(6) 2021.
 
Additional Manufacturer Narrative
The device was not available for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, it is noted the patient had pain, a dislocation and subsequent manipulation under anesthesia after a bending-twisting motion.It cannot be concluded the reported events were associated with a mal performance of the implant.The patient had a successful reduction and was instructed to follow-up with her orthopedic doctor; however, follow-up records were not provided.No further clinical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for hip systems revealed that dislocation has been identified as a warnings and precautions, that may result from improper selection of the neck length and cup, stem positioning, muscle looseness and/or malpositioning of components.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, limited range of motion, patient anatomy or abnormal loading of limb.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXINIUM FEM HD 12/14 36 MM L/+8
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15136398
Report Number1020279-2022-03546
Device Sequence Number14751530
Product Code JDI
UDI-Device Identifier03596010477293
UDI-Public03596010477293
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K022958
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71343608
Device Catalogue Number71343608
Device Lot Number19HM01208
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2022
Supplement Date Received by Manufacturer09/16/2022
Initial Report FDA Received Date07/29/2022
Supplement Report FDA Received Date09/21/2022
Date Device Manufactured08/01/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PN: 71335552 / LOT: 17BM08020A; PN: 71356006 / LOT: 16FM14196
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age77 YR
Patient SexFemale
-
-