Model Number 106524US
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Bacterial Infection (1735); Post Operative Wound Infection (2446); Skin Infection (4544)
Event Date 07/06/2022
Event Type  Injury  
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
 
Event Description
It was reported that the patient was admitted for infection on (b)(6) 2022.The patient was listed status 1a due to persistent infection.The patient had persistent drainage from the chest wound and development of hypergranular tissue at the driveline exit site.The driveline tract was infected from exit site to mediastinum.The drainage was positive for mycobacterium abscesses and was treated with intravenous antibiotics and on (b)(6) 2022 the patient underwent heart transplantation.The device otherwise operated as expected.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the pump was explanted.A correlation between the device and the report of infection could not be conclusively determined.Infection is listed in the instructions for use as a potential adverse event that may be associated with the use of heartmate 3 left ventricular assist system.Infection has been previously investigated and will continue to be monitored through quality data reviews, which are conducted on production and post product signals to evaluate if products are conforming to product requirements.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE 3 LVAS IMPLANT KIT
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key15136536
MDR Text Key296950254
Report Number2916596-2022-12211
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00813024013297
UDI-Public00813024013297
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date11/18/2022
Device Model Number106524US
Device Catalogue Number106524US
Device Lot Number7787128
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/08/2022
Initial Date FDA Received07/29/2022
Supplement Dates Manufacturer Received08/10/2022
Supplement Dates FDA Received08/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age15 YR
Patient SexMale
Patient Weight116 KG
Patient EthnicityHispanic
-
-