BECTON DICKINSON MEDICAL SYSTEMS BD LUER-LOK¿ SYRINGE WITH ATTACHED NEEDLE 23 G X 1 1/2 IN.; PISTON SYRINGE Back to Search Results
Catalog Number 309589
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while using a bd luer-lok¿ syringe with attached needle 23 g x 1 1/2 in., leakage occurred.The following information was provided by the initial reporter: when the diluent was injected into the vial, the luerlock of the syringe started spraying diluent and an unknown amount of diluent went into the vaccine vial.
 
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Fda notified: the initial reporter also notified the fda via medwatch # mw5110482.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that while using a bd luer-lok¿ syringe with attached needle 23 g x 1 1/2 in., leakage occurred.The following information was provided by the initial reporter: when the diluent was injected into the vial, the luerlock of the syringe started spraying diluent and an unknown amount of diluent went into the vaccine vial.
 
Additional Manufacturer Narrative
H.6.Investigation summary: since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident.No root cause can be determined as no samples were received.Furthermore, a device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Examination of the product involved may provide clarification as to the cause for the reported failure.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD LUER-LOK¿ SYRINGE WITH ATTACHED NEEDLE 23 G X 1 1/2 IN.
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15136552
Report Number1213809-2022-00443
Device Sequence Number14761017
Product Code FMF
UDI-Device Identifier00382903095896
UDI-Public00382903095896
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K110771
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/31/2025
Device Catalogue Number309589
Device Lot Number0308897
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2022
Supplement Date Received by Manufacturer07/30/2022
Initial Report FDA Received Date07/29/2022
Supplement Report FDA Received Date08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-