| Model Number |
B1070-060 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the procedure was performed to treat the left posterior tibial artery with moderate stenosis.An armada 35 7mm/60mm was advanced to the target lesion without issue; however, on the first inflation, the balloon ruptured at 6 atmospheres.Therefore, the device was removed and replaced.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection was performed on the returned device.The reported balloon rupture was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported compliant.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
|
| |
|