SMITH & NEPHEW, INC. GNS II RESURF PAT 26MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Model Number |
71420580 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Joint Laxity (4526)
|
| Date of Event |
07/13/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, after a tka had been performed on (b)(6) 2010 with a genesis ii system, the patient experienced instability in the knee for a couple of months.This adverse event was addressed via revision surgery on (b)(6) 2022.Upon inspection, it was noticed that the gns ii resurf pat 26mm was loose.Patient current health status is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device did not confirm the stated failure mode.The clinical/medical evaluation concluded that based on a review of the two undated, unlabeled photos of the explant, two of the chart sticks, and two undated, unlabeled x-ray photos were provided but do not indicate a root cause of the broken insert post.The revision finding of the reported loosening was confirmed to be related to the broken post.However, the patient¿s body habitus of 238 pounds could not be ruled out as a possible contributory factor.Per the ifu for the knee systems; ¿the implant can break or become damaged because of strenuous activity or trauma and has a finite expected service life and may need to be replaced in the future.¿ it should also be noted the implant was implanted 12 years prior to the revision.According to the report, the patient¿s current health status is unknown.Therefore, the patient impact beyond the instability, the revision, and expected transient post-op convalescence period cannot be determined.No further clinical assessment is warranted at this time.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.The contribution of the device to the reported event could not be corroborated.Factors and/or potential causes that could contribute to the reported event have been identified in the risk management file.This event was evaluated by our quality team and the evaluation could not confirm this failure as a result of a non conforming product, as the device was found to be too damaged from prolonged use.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device was returned and evaluated.A visual inspection of the returned device did not confirm the stated failure mode.The clinical/medical evaluation concluded that based on a review of the two undated, unlabeled photos of the explant, two of the chart sticks, and two undated, unlabeled x-ray photos were provided but do not indicate a root cause of the broken insert post.The revision finding of the reported loosening was confirmed to be related to the broken post.However, the patient¿s body habitus of 238 pounds could not be ruled out as a possible contributory factor.Per the instruction for use for the knee systems; ¿the implant can break or become damaged because of strenuous activity or trauma and has a finite expected service life and may need to be replaced in the future.¿ it should also be noted the implant was implanted 12 years prior to the revision.According to the report, the patient¿s current health status is unknown.Therefore, the patient impact beyond the instability, the revision, and expected transient post-op convalescence period cannot be determined.No further clinical assessment is warranted at this time.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.The contribution of the device to the reported event could not be corroborated.Factors that could contribute to the reported event include traumatic injury, size selected, postoperative care.This event was evaluated by our quality team and the evaluation could not confirm this failure as a result of a non conforming product, as the device was found to be too damaged from prolonged use.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|