• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GENESIS PIN AND DRILL SET; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GENESIS PIN AND DRILL SET; PLATE, FIXATION, BONE Back to Search Results
Model Number 114968
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 07/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during a tka surgery, a genesis pin and drill set broke when the surgeon was drilling the holes, prior to pinning to secure and finish the tibial ream.The broken part could not be removed and remained in the patient's tibia.Surgery was finished without further usage of the drill.No significant delays reported.Current health status of patient is unknown.
 
Additional Manufacturer Narrative
Section h3, h6: the device was not returned for evaluation but the pictures were reviewed, and revealed that the device is broken.The clinical/medical investigation concluded that, this case reports the breakage of the genesis pin and drill set, intra-operatively the broken part was left inside of patients tibial.No clinically relevant supporting documentation was provided for inclusion in this medical investigation.Therefore, the root cause of the reported breakage could not be determined.The genesis pin and drill sets are comprised of ss, and they are manufactured and intended as externally communicating devices and are not approved for long term internal tissue exposure and long-term implantation data is not available.Since the broken part was retained the patient¿s tibial micro-motion and/or migration is unlikely.However, we cannot make conclusions on the impact of the non-implantable foreign body.Based on the information provided, the surgery was completed without further usage of the drill with an approximately 30-minute surgical delay.The impact to the patient beyond that which has already been reported cannot be confirmed nor concluded due to the lack of information.Should any additional medical information be provided, this case would be re-assessed.Therefore, no further clinical/medical investigation is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries, therefore damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported incident could be corroborated as the picture reveal that the device is broken.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GENESIS PIN AND DRILL SET
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15136635
Report Number1020279-2022-03554
Device Sequence Number20150697
Product Code HRS
UDI-Device Identifier03596010051332
UDI-Public03596010051332
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K123055
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number114968
Device Catalogue Number114968
Device Lot Number20GM18497
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2022
Supplement Date Received by Manufacturer09/05/2022
Initial Report FDA Received Date07/29/2022
Supplement Report FDA Received Date09/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/29/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-