Internal complaint reference (b)(4).H10 h3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A complaint history review found no similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A visual inspection of the returned device found that it is not in its original packaging.The distal end of the device is fractured and there is debris inside the device and in the crack.The proximal end of the device is whole and there is some debris visible inside the device.Based on the condition of the product material found during visual inspection, additional material testing is not required.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.Internal complaint reference (b)(4).A1: patient identifier space updated from "none" to blank since we dont have information about the patient but know there was one patient involved.
|