• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. EVOS SMALL 2.5MM DRILL W/AO QC LONG; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. EVOS SMALL 2.5MM DRILL W/AO QC LONG; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Model Number 71175022
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/19/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case(b)(4).
 
Event or Problem Description
It was reported that, during a fracture of the tibial plateau procedure, when drilling the bone, the first evos small 2.5mm drill w/ao qc long broke in the patient and the second evos small 2.5mm drill w/ao qc long was bent.Incident occurred with instruments inside the patient.The procedure was completed, without delay, using a s+n back-up device.Patient was not harmed.
 
Event or Problem Description
It was reported that, during a fracture of the tibial plateau procedure, when drilling the bone, an evos small 2.5mm drill w/ao qc long broke in the patient.The fractured piece was left inside of the patient.The procedure was completed, without delay, using a s+n back-up device.Patient's current health status is unknown.
 
Additional Manufacturer Narrative
The device was not returned for evaluation but the pictures were reviewed, and revealed that the tip of the device is broken.The clinical/medical investigation concluded that, during a tibial plateau fracture procedure, when drilling the bone, the first evos small 2.5mm drill w/ao qc long broke in the patient.The fractured piece was left inside of the patient.The chemical makeup of the drill bit is the common 17.4 medical stainless steel and is not intended for long term implantation.As of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.The patient's current condition is unknown, and the patient impact could not be determined based on the limited information provided.Should any additional clinical information be provided, this complaint will be re-evaluated.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.As the device can no longer fit its purpose, the contribution of the device to the reported event could be corroborated.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Additional Manufacturer Narrative
The associated device was returned and evaluated.The visual inspection revealed the tip of the device broke off.The piece was not returned.The device shows signs of significant wear and use.The clinical/medical investigation concluded that, during a tibial plateau fracture procedure, when drilling the bone, the first evos small 2.5mm drill w/ao qc long broke in the patient.The fractured piece was left inside of the patient.The chemical makeup of the drill bit is the common 17.4 medical stainless steel and is not intended for long term implantation.As of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.The patient's current condition is unknown, and the patient impact could not be determined based on the limited information provided.Should any additional clinical information be provided, this complaint will be re-evaluated.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for care, maintenance, cleaning, and sterilization of smith & nephew orthopedics devices for single-use devices revealed that as the devices are sold both nonsterile and sterile and often removed from their original packaging to be placed in a containment device they should be cleaned and/or inspected prior to sterilization.The device should not be reprocessed or reused if comes in contact with blood, tissue or bodily fluids.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.With the results of this investigation the root cause of this event could not be determined.This is a single use device, surgical technique or damage from misuse are likely potential factors that could contribute to the reported event.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS SMALL 2.5MM DRILL W/AO QC LONG
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15136782
Report Number1020279-2022-03565
Device Sequence Number20182155
Product Code HWE
UDI-Device Identifier00885556635766
UDI-Public00885556635766
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71175022
Device Catalogue Number71175022
Device Lot Number20FUQ0003C
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/19/2022
Supplement Date Received by Manufacturer07/19/2022
10/26/2022
03/08/2023
Initial Report FDA Received Date07/30/2022
Supplement Report FDA Received Date10/04/2022
10/27/2022
03/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/31/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age50 YR
Patient SexFemale
-
-