BOSTON SCIENTIFIC CORPORATION CRE PRO; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00558630
Medical Device Problem Codes Inflation Problem (1310); Material Rupture (1546); Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a cre pro wireguided dilatation balloon was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, it was noted that the balloon could not be inflated.The procedure was completed with another cre pro wireguided dilatation balloon.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.This event has been deemed reportable based on the investigation results of balloon burst.
 
Additional Manufacturer Narrative
The patient's exact weight was reported to be 49.5 kg.(b)(4).Investigation results: the returned cre pro wireguided dilatation balloon was analyzed and a visual and microscopic examination found that the balloon was ruptured.No damages were found to the catheter of the device and the guidewire was noted to be bent.With all the available information, boston scientific concludes the reported event was not confirmed.The rupture found in the balloon is likely to have occurred due to factors encountered during the procedure; excess pressure, interaction with other devices, or anatomical affairs.It is possible that factors encountered during the procedure, such as the interaction with any surface, could create friction on the balloon and cause the problem of failure to inflate during the procedure.Therefore, the most probable root cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.A labeling review was performed and from the information available, this device was used per the instructions for use (ifu)/product label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE PRO
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15136892
Report Number3005099803-2022-04087
Device Sequence Number7372622
Product Code FGE
UDI-Device Identifier08714729797517
UDI-Public08714729797517
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K112994
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/13/2021
Device Model NumberM00558630
Device Catalogue Number5863
Device Lot Number0022935377
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/06/2022
Initial Report FDA Received Date07/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/13/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age71 YR
Patient SexFemale
Patient Weight50 KG
-
-