EXACTECH, INC. NV GXL LINER LIPPED 32MM ID, GROUP 1 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Model Number |
NV EHXL ALIP LNR G1 32MM |
| Medical Device Problem Codes |
Fracture (1260); Naturally Worn (2988); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Ambulation or Postural Difficulties (2544); Metal Related Pathology (4530)
|
| Date of Event |
06/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, this patient had a tha on (b)(6) 2017.On (b)(6) 2022, the patient underwent a hip revision.The patient left in stable condition.No other known info.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending evaluation.Biolox delta femoral head, 32mm od, -3.5mm - cn: 170-32-93 / sn: 4413043.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the reason for the revison was not reported but is most likely the result of prosthesis wear and/or fracture of the actetabular liner leading to articulation between the femoral head and acetabular cup.The contributions of implant positioning and/or patient related factors to the sequence of events cannot be determined from the reported information; however, the combination of risk factors may have also been a contributing factor to the early prosthesis wear/fracture.
|
| |
|
Additional Manufacturer Narrative
|
|
H7: z-1732-2022.
|
| |
|
Event or Problem Description
|
|
Additional information received via legal filing: the patient experienced complications, including but not limited to severe pain, difficulty in walking, metallosis and fractured liner.The patient then underwent a revision surgery on (b)(6) 2022, approximately 4 years, 8 months, after initial implant.It is believed that complications were due to premature polyethylene wear.There is no other patient demographic or medical history available.There is no device return.No additional information is available.
|
| |
|
Event or Problem Description
|
|
As reported, this patient had a tha.The patient experienced complications, including but not limited to severe pain, difficulty in walking, metallosis, and a fractured liner.The patient then underwent a revision surgery, approximately 4 years, 8 months, after initial implant.There is no other patient demographic or medical history available.There is no device return.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
D10: concomitants: 2609960 132-32-51 - nv gxl liner lipped 32mm id, group 1 cups.2972181 101-45-30 - 4.5mm drill bit 30mm.4413043 170-32-93 - biolox delta femoral head 32mm od, -3.5mm.4782632 186-01-50 - integrip cc, cluster 50mm, g1.4922785 180-65-30 - alteon 6.5mm screw, 30mm.4931791 188-01-04 - wedge plasma x/o sz 4.4964403 180-65-25 - alteon 6.5mm screw, 25mm." this follow-up report is being submitted to relay additional and/or corrected information.The following sections were added/corrected: b1, b2, b5, d1/d2a/d2b, d4, d10, h4, h6: clinical code, impact code, medical device problem code, component code, type of investigation, investigation findings, investigation conclusion.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: h6: investigation findings and investigation conclusions.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|