| Model Number |
3288 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.
|
| |
|
Event or Problem Description
|
|
It was reported the patient experienced ineffective therapy.Lead diagnostics indicated high impedances on all contacts.Reprogramming did not resolve the issue.X-ray confirmed lead fracture and the leads are tangled in the ipg pocket site.In turn, surgical intervention may occur to address the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated the patient underwent surgical intervention wherein the lead was explanted and replaced to address the issue.
|
| |
|