| Model Number |
5532-G-711 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
07/11/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.The following devices were also listed in this report: triathlon ps fem comp #7r-cem; cat#5515-f-702; lot#aeb2d triathlon prim cem fxd bplt #7; cat#5520b700; lot#aa44ee triathlon asymmetric x3 patella; cat#5551-g-381; lot#0k0y it cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
|
| |
|
Event or Problem Description
|
|
It was reported that a stage 1 revision was performed on the patient's right knee due to infection.All components were removed and a spacer was placed.
|
| |
|
Event or Problem Description
|
|
It was reported that a stage 1 revision was performed on the patient's right knee due to infection.All components were removed and a spacer was placed.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding infection involving a triathlon insert was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot or sterile lot referenced.Conclusions: all stryker products sold as sterile are validated to a minimum sterility assurance level sal of 10^-6 in accordance with applicable iso standards.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.Lot code corrected for the below device, triathlon prim cem fxd bplt #7; cat#5520b700; lot#aa44e.
|
| |
|