• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. TRIAL, LARGE, PEGGED GLENOID; TEMPLATE FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. TRIAL, LARGE, PEGGED GLENOID; TEMPLATE FOR CLINICAL USE Back to Search Results
Model Number TRIAL, LARGE, PEGGED GLENOID
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2022 it was reported by a sales representative via sems that both an ar-9236-03pp glenoid trial and an ar-9236-03 pegged glenoid trial broke immediately upon pushing into the glenoid with hand.This occurred during a total eclipse total shoulder procedure.All was retrieved from the patient.The procedure was completed successfully without a trial being used.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Complaint confirmed.Upon visual inspection, it was noted that the central post was broken.The broken piece was note returned for evaluation.The cause remains undetermined, however a probable cause can be attributed to damage incurred during the insertion and/or removal process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIAL, LARGE, PEGGED GLENOID
Common Device Name
TEMPLATE FOR CLINICAL USE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15137873
Report Number1220246-2022-05328
Device Sequence Number14856549
Product Code HWT
UDI-Device Identifier00888867059498
UDI-Public00888867059498
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTRIAL, LARGE, PEGGED GLENOID
Device Catalogue NumberAR-9236-03
Device Lot Number113601526
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2022
Supplement Date Received by Manufacturer07/14/2022
Initial Report FDA Received Date08/01/2022
Supplement Report FDA Received Date10/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/30/2015
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1