• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH CMN FEMORAL NAIL, CCD 130, RIGHT¸ 11.5 MM, 30 CM; HIP PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH CMN FEMORAL NAIL, CCD 130, RIGHT¸ 11.5 MM, 30 CM; HIP PROSTHESIS Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent an initial right hip arthroplasty and approximately six (6) weeks post-implantation started to experience pain.Implant fracture was discovered in post-op x-rays.Revision is not planned, the patient's condition will be continued to be monitored.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: ball nose guide wire 100cm item# 281001100 lot# 508380.Znn, cmn lag screw, 10.5 mm, 105 mm including set screw item# 47-2485-105-10 lot# 3075443.5.0 mm diameter cortical screw - red fixed angle 3.5 mm hex head item# 47-2484-035-50 lot# 65225644.5.0 mm diameter cortical screw - red fixed angle 3.5 mm hex head item# 47-2484-032-50 lot# 65167613.Report source¿ foreign ¿ france.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0009613350 - 2022 - 00399.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: b4, b5, g3, g6, h1, h2, h3, h6, h10.No product was returned, or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Device are used for treatment.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: first review: x-ray: components in place, fracture not healed.No exact date of occurrence, proximately 6 weeks post op.Second review: good healing, no pain, ambulating with partial support.X-ray; no secondary displacement with nail in place.Third review: slight pain, normal finding.X-ray; fracture of nail.Options presented and patient does not want a revision, will continue to ambulate with support.Heel cup prescribed; patient refused.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CMN FEMORAL NAIL, CCD 130, RIGHT¸ 11.5 MM, 30 CM
Common Device Name
HIP PROSTHESIS
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key15138411
Report Number0009613350-2022-00398
Device Sequence Number14887100
Product Code HSB
UDI-Device Identifier00889024299146
UDI-Public(01)00889024299146(17)300131(10)3025294
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number47-2493-302-11
Device Lot Number3025294
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2022
Supplement Date Received by Manufacturer10/05/2022
Initial Report FDA Received Date08/01/2022
Supplement Report FDA Received Date10/17/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/20/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Other;
Patient Age76 YR
Patient SexFemale
-
-