(b)(4).Concomitant medical products: ball nose guide wire 100cm item# 281001100 lot# 508380.Znn, cmn lag screw, 10.5 mm, 105 mm including set screw item# 47-2485-105-10 lot# 3075443.5.0 mm diameter cortical screw - red fixed angle 3.5 mm hex head item# 47-2484-035-50 lot# 65225644.5.0 mm diameter cortical screw - red fixed angle 3.5 mm hex head item# 47-2484-032-50 lot# 65167613.Report source¿ foreign ¿ france.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0009613350 - 2022 - 00399.
|
|
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: b4, b5, g3, g6, h1, h2, h3, h6, h10.No product was returned, or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Device are used for treatment.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: first review: x-ray: components in place, fracture not healed.No exact date of occurrence, proximately 6 weeks post op.Second review: good healing, no pain, ambulating with partial support.X-ray; no secondary displacement with nail in place.Third review: slight pain, normal finding.X-ray; fracture of nail.Options presented and patient does not want a revision, will continue to ambulate with support.Heel cup prescribed; patient refused.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|