MAUDE Adverse Event Report: TELEFLEX MEDICAL PE ADULT-PED WET LF 6/CS; BOTTLE, COLLECTION, VACUUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL PE ADULT-PED WET LF 6/CS; BOTTLE, COLLECTION, VACUUM Back to Search Results
Model Number IPN028446
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
We have had 3 cases in the last week in which we have identified a defect in the pleur-evac, no lot numbers or defective product a vailable.The is a defect at the junction of the rubber tubing to the corrugated tubing (red to blue connection).We have had to change out these 3 vacs for these patients within 24 hours post op.We have also completed fda form 3500.
 
Additional Manufacturer Narrative
Qn#(b)(4).The device has not been returned for investigation.
 
Additional Manufacturer Narrative
Qn# (b)(4).The device history record of batch number 74m2101787 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non-conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled & inspected according to our specifications.One unit of catalog number a-7000-08lf was received for analysis.The ats connector red & blue were received detached.The o-ring was visually inspected , and no damages were observed.The patient tubing with the ats connectors were connected to a source of air and was immerse under water.No bubbling's were observed that can confirmed a leak on ats connector.In summary, no corrective action can be established since no problem found on sample.Customer complaint cannot be confirmed since no problem found on sample.
 
Event or Problem Description
We have had 3 cases in the last week in which we have identified a defect in the pleur-evac, no lot numbers or defective product a vailable.The is a defect at the junction of the rubber tubing to the corrugated tubing (red to blue connection).We have had to change out these 3 vacs for these patients within 24 hours post op.We have also completed fda form 3500.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PE ADULT-PED WET LF 6/CS
Common Device Name
BOTTLE, COLLECTION, VACUUM
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key15138460
Report Number3004365956-2022-00053
Device Sequence Number12173842
Product Code KDQ
UDI-Device Identifier34026704631279
UDI-Public34026704631279
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/16/2024
Device Model NumberIPN028446
Device Catalogue NumberA-7000-08LF
Device Lot Number74M2101787
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2022
Supplement Date Received by Manufacturer09/23/2022
Initial Report FDA Received Date08/01/2022
Supplement Report FDA Received Date09/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-