| Model Number |
IPN028446 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
We have had 3 cases in the last week in which we have identified a defect in the pleur-evac, no lot numbers or defective product a vailable.The is a defect at the junction of the rubber tubing to the corrugated tubing (red to blue connection).We have had to change out these 3 vacs for these patients within 24 hours post op.We have also completed fda form 3500.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The device has not been returned for investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The device history record of batch number 74m2101787 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non-conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled & inspected according to our specifications.One unit of catalog number a-7000-08lf was received for analysis.The ats connector red & blue were received detached.The o-ring was visually inspected , and no damages were observed.The patient tubing with the ats connectors were connected to a source of air and was immerse under water.No bubbling's were observed that can confirmed a leak on ats connector.In summary, no corrective action can be established since no problem found on sample.Customer complaint cannot be confirmed since no problem found on sample.
|
| |
|
Event or Problem Description
|
|
We have had 3 cases in the last week in which we have identified a defect in the pleur-evac, no lot numbers or defective product a vailable.The is a defect at the junction of the rubber tubing to the corrugated tubing (red to blue connection).We have had to change out these 3 vacs for these patients within 24 hours post op.We have also completed fda form 3500.
|
| |
|