| Model Number |
N/A |
| Medical Device Problem Codes |
Thermal Decomposition of Device (1071); Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is complete, a follow up/final report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the unit was not powering on and the inlet module is burnt.The issue did not occur during surgery.No adverse events were reported as a result of this malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined that the unit was not powering on and the fuse box on the power inlet module was melted.The power inlet module was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional event information available.
|
| |
|
Search Alerts/Recalls
|