| Model Number |
7097C |
| Medical Device Problem Code |
High Readings (2459)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's weight was not provided.
|
| |
|
Event or Problem Description
|
|
The customer reported that he obtained blood glucose readings of 357 mg/dl at 9:27 a.M.And 83 mg/dl at 9:29 a.M.With the contour meter.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.Replacement meter kit and test strips were sent to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer returned a contour meter (serial # (b)(6)) and contour test strips (lot # 1aj3b01d) for evaluation.The serial number of the returned meter and lot number of the returned test strips were different from the one associated with event.An in-house testing was performed using the blood sample with the returned meter and test strips as well as with the in-house test strips from lot # 1ej3b01d, which is same as the lot # associated with the event and it gave a satisfactory performance.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record was reviewed for the suspected contour meter and no manufacturing anomalies were found.
|
| |
|