ASCENSIA DIABETES CARE US INC. CONTOUR; BLOOD GLUCOSE TEST STRIPS Back to Search Results
Model Number 7097C
Medical Device Problem Code High Readings (2459)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's weight was not provided.
 
Event or Problem Description
The customer reported that he obtained blood glucose readings of 357 mg/dl at 9:27 a.M.And 83 mg/dl at 9:29 a.M.With the contour meter.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.Replacement meter kit and test strips were sent to the customer.
 
Additional Manufacturer Narrative
The customer returned a contour meter (serial # (b)(6)) and contour test strips (lot # 1aj3b01d) for evaluation.The serial number of the returned meter and lot number of the returned test strips were different from the one associated with event.An in-house testing was performed using the blood sample with the returned meter and test strips as well as with the in-house test strips from lot # 1ej3b01d, which is same as the lot # associated with the event and it gave a satisfactory performance.
 
Additional Manufacturer Narrative
A device history record was reviewed for the suspected contour meter and no manufacturing anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR
Common Device Name
BLOOD GLUCOSE TEST STRIPS
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
100 summit lake drive
valhalla NY 10595
Manufacturer (Section G)
PHC CORPORATION (REGISTRATION NUMBER: 1000134386)
2131-1 minamigata
toon ehime, 791-0 395
JA   791-0395
Manufacturer Contact
shweta gulati
100 summit lake drive
valhalla, NY 10595
9142361830
MDR Report Key15138968
Report Number1810909-2022-00136
Device Sequence Number12165954
Product Code NBW
UDI-Device Identifier10301937097505
UDI-Public10301937097505
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K062058
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/31/2023
Device Model Number7097C
Device Catalogue Number7097C
Device Lot Number1EJ3B01D
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/14/2022
Supplement Date Received by Manufacturer10/27/2022
11/25/2022
Initial Report FDA Received Date08/01/2022
Supplement Report FDA Received Date11/18/2022
11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age82 YR
Patient SexMale
-
-