• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO MISSION (US) 2.5/22; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO MISSION (US) 2.5/22; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 453950
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
An orsiro mission drug-eluting stent system was selected for treatment of a severely calcified lesion in the mid lad.The stent dislodged while attempting to advance the device to the lesion which was heavily calcified and had likely concurrent intimal disruption at the distal edge of a more proximal orsiro mission stent.
 
Additional Manufacturer Narrative
Combination product: yes.Neither the complaint instrument nor the angiographic material was returned.Therefore, no technical investigation on the subject could be performed.The product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.Considering the physicians event description, the most probable root cause for the complaint event is related to the patients anatomy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO MISSION (US) 2.5/22
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key15139280
Report Number1028232-2022-03774
Device Sequence Number5646809
Product Code NIQ
UDI-Device Identifier07640130456068
UDI-Public07640130456068
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/24/2024
Device Model Number453950
Device Catalogue NumberSEE MODEL NO.
Device Lot Number01221283
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2022
Initial Report FDA Received Date08/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
-
-