| Model Number |
8888145058P |
| Medical Device Problem Codes |
Material Puncture/Hole (1504); Defective Component (2292)
|
| Health Effect - Clinical Codes |
Air Embolism (1697); Hemorrhage/Blood Loss/Bleeding (1888); Hemorrhagic Stroke (4417); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
06/26/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, when the patient was sleeping and woke up, it was noted that the cvc (central venous catheter) gauze was soaked in blood.The patient was presented to the er (emergency room) where the cvc was assessed by the nephrologist.It was noted to be oozing fresh blood from the venous lumen, just prior to its connections at the hub where the lumen width increases.The nephrologist stated whenever the patient moved, or stood up, there would be fresh blood leaking from what appears to be a crack.The patient was not on dialysis at the time of the incident so the lumens were locked.There was a blood loss of approximately 30ml but no blood transfusion required.Flushing was done pre and post treatment.The catheter was not repaired and no luer adapter issue.While still functional in the patient, the cvc exit site was cleansed and tego's with chg (chlorhexidine gluconate) were utilized.Clamps are rotated up and down the lumen each treatment.Once removed it was wiped with a cavi wipe.No ointment or cleaning agent used on the catheter or the surrounding skin.No other products being utilized with the device.Nothing unusual observed on the device prior to use.Cvc was double clamped, patient was admitted to the hospital for 4 days until the cvc could be safely removed and missed the treatment.A new competitor's cvc was placed as the catheter was not safe once the lumen was cracked.The replacement catheter was successfully used and completed the procedure.There were no other patient symptoms or complications associated with the event.
|
| |
|
Event or Problem Description
|
|
According to the reporter, when the patient was sleeping and woke up, it was noted that the cvc (central venous catheter) gauze was soaked in blood.The patient was presented to the er (emergency room) where the cvc was assessed by the nephrologist.It was noted to be oozing fresh blood from the venous lumen, just prior to its connections at the hub (bifurcate) where the lumen width increases (bulged - exactly where the leak occurred).The nephrologist stated whenever the patient moved, or stood up, there would be fresh blood leaking from what appears to be a crack.The patient was not on dialysis at the time of the incident so the lumens were locked.There was a blood loss of approximately 30ml but no blood transfusion required.Flushing was done pre and post treatment.The catheter was not repaired and no luer adapter issue.While still functional in the patient, the cvc exit site was cleansed and tego's w ith chg (chlorhexidine gluconate) were utilized.Clamps are rotated up and down the lumen each treatment.Once removed it was wiped with a cavi wipe.No ointment or cleaning agent used on the catheter or the surrounding skin.No other products being utilized with the device.Nothing unusual observed on the device prior to use.Cvc was double clamped below the leak, patient was admitted to the hospital for 4 days until the cvc could be safely removed and missed the treatment.A new competitor's cvc was placed as the catheter was not safe once the lumen was cracked.The replacement catheter was successfully used and completed the procedure.There were no other patient symptoms or complications associated with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: a3, b5, g1(manufacturer name), g3 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: patient codes - e2402 (permanent damage).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, when the patient was sleeping and woke up, it was noted that the cvc (central venous catheter) gauze was soaked in blood (had hemorrhage/bleeding).The patient was presented to the er (emergency room) where the cvc was assessed by the nephrologist.It was noted to be oozing fresh blood from the venous lumen, just prior to its connections at the hub (bifurcate) where the lumen width increases (bulged - exactly where the leak occurred).It was also stated that the doctor came and seen that there was a hole in the line with air bubbles.It had also fluid leak.The nephrologist stated whenever the patient moved, or stood up, there would be fresh blood leaking from what appears to be a crack.The patient was not on dialysis at the time of the incident so the lumens were locked.There was a blood loss of approximately 30ml but no blood transfusion required.Flushing was done pre and post treatment.The catheter was not repaired and no luer adapter issue.While still functional in the patient, the cvc exit site was cleansed and tego's with chg (chlorhexidine gluconate) were utilized.Clamps are rotated up and down the lumen each treatment.Once removed it was wiped with a cavi wipe.No ointment or cleaning agent used on the catheter or the surrounding skin.No other products being utilized with the device.Nothing unusual observed on the device prior to use.Cvc was double clamped below the leak, the doctor flushed and clamped it to come back the next day for an operation.The patient was admitted to the hospital for 4 days until the cvc could be safely removed and missed the treatment.A new competitor's cvc was placed as the catheter was not safe once the lumen was cracked.The replacement catheter was successfully used and completed the procedure.The patient went to leave and had air embolism and stroke/cerebrovascular accident.It was a second line in 6 weeks with a hole in it.It was also stated that the outcomes attributed to adverse event were hospitalization, disability or permanent damage.There were no other patient symptoms or complications associated with the event.
|
| |
|