| Model Number |
0P9PTVT500 |
| Health Effect - Clinical Code |
Unspecified Tissue Injury (4559)
|
| Date of Event |
06/06/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The therapy wrap (foot/ankle wrap) used by the patient was not returned with the device, so we used a new version from inventory of the same model to simulate therapy performed.The prescription provided by the dme did not indicate the modalities that the device and wrap were to be used for therapy.It only indicated to use for 28 days (no temperature or on/off times).The patient agreement was not signed to show that the patient had been instructed in the correct use of the vascutherm unit.The customer requested an rma for the device for leaking; it was after the device had been returned and begun evaluation that the customer called us back and reported the pt issue.We had to retrieve device from the repair dept.And quarentine for evaluation related to suspect pt incident.Device performed as intended during preliminary investigation; further testing requested by engineering.Device is quarentined in quality assurance until further testing can be performed.A follow up report will be made after further testing is completed.
|
| |
|
Event or Problem Description
|
|
After two days of use, pt complained device was leaking and there is an issue with her foot and hopes its not caused from the vt5.On (b)(6), dme replaced unit with another and no longer experienced any further issues.Dr discontinued use on (b)(6).Even was not reported to manufacturer until (b)(6) 2022.
|
| |
|