Model Number 04.038.200
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Pain (1994)
Date of Event 07/05/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2022, patient came to er stating that she had a ground level fall and was complaining of increase left hip pain.Patient had original trochanteric fixation nail-advanced (tfna) implanted on (b)(6) 2022.Surgeon scheduled a revision surgery on (b)(6) 2022 for removal of broken hardware which was confirmed by er x-rays on (b)(6) 2022.Hardware was removed and a trochanteric fixation nai (tfn) was chosen for the repair because the tfn had a larger diameter proximal body and lag screw.Implants were removed and documented and tfn was implanted with success.It was described that implant failed at proximal end of nail, proximal to lag screw insertion hole.Patient experienced an adverse event.The surgery was successfully completed.This report is for one (1) tfna fenestrated screw 100mm.This is report 2 of 2 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is not expected to be returned for manufacturer review/investigation.The photo was returned to depuy synthes for evaluation.Visual analysis of the photo revealed that the tfna fenestrated screw 100mm had no signs of issue or defect.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed for the tfna fenestrated screw 100mm.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.A device history record (dhr) review was conducted: part number: 04.038m200sp.Lot number: 705p376.Part manufacture date: 07-mar-2022.Manufacturing location: (b)(4).Part expiration date: n/a.Nonconformance noted: n/a.Dhr record review: a review of the device history record revealed no complaint related anomalies.The device history record shows this lot of tfna fenestrated screw 100mm - sterile product was processed through the normal manufacturing and inspection operations with no rework nor nonconformities noted.This lot met all dimensional, visual, and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.A review of the raw material device history record revealed this lot met all specifications with no nonconformance noted.This raw material lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.Elmira ships this product to monument as part number 04.038m200sp.Monument performs the sterile processing of this part changing the part number to 04.038.200s.Monument will need to perform the dhr review for the monument operations.This lot met all dimensional, visual, and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TFNA FENESTRATED SCREW 100MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
ELMIRA
35 airport road
horseheads NY 14845
Manufacturer Contact
kate karberg
1302 wrights lane east
west chester, PA 19380
8472871282
MDR Report Key15140463
Report Number2939274-2022-03048
Device Sequence Number7372600
Product Code HSB
UDI-Device Identifier10886982098764
UDI-Public(01)10886982098764
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K160167
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.038.200
Device Catalogue Number04.038.200
Device Lot Number705P376
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/05/2022
Initial Report FDA Received Date08/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
14/130 DEG TI CANN TFNA 400/LEFT-STERILE
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexFemale