| Model Number |
442960 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device manufacture date: unknown.Medical device type: additional device type: pch.Medical device expiration date: unknown.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd max¿ enteric parasite panel there was discrepant results.No patient impact.The following information was provided by the initial reporter: customer reported discrepant results on 7/8, giardia was positive and on 7/10 repeated duplicates were negative.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields have been corrected with the following information: b.5 describe event or problem: bd max¿ instrument, remanufactured.D.1 medical device brand name: bd max¿ instrument, remanufactured.D.2a: common device name: instrumentation for clinical multiplex test systems.D.4: medical device catalog #: 441927.D.4 unique identifier (udi) #: (b)(4).D.4 medical device serial #: (b)(6).G.5 pma/510(k)#: k111860, k120138.H.4: device manufacture date: 03/05/2019.H.5: labeled for single use: no.
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd max¿ instrument, remanufactured there was discrepant results.No patient impact.The following information was provided by the initial reporter: customer reported discrepant results on 7/8, giardia was positive and on 7/10 repeated duplicates were negative.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6 investigation summary: the complaint alleges the bd max instrument (catalog number 441927 and serial number (b)(6)) had "discrepant results." service retrieved the customer instrument database for further analysis.After database review, sme recommended renormalizing of both readers and to perform an instrument health check.Fse was dispatched and they normalized both instrument readers and performed the necessary instrument and reader tests and checks.Qualification run was performed, and the instrument passed.This complaint is confirmed by service.The root cause cannot be determined with the information provided.Samples were not received by quality for investigation and thus, returned material investigation could not occur.If samples are received at a later date, the complaint may be reopened.Review of device history record for instrument (b)(6) is not required because this complaint does not allege an early life failure or failure at installation and has changed configuration since release from manufacturing due to service repairs/pms.Device was installed on 03jun2019.Service history review was performed for the instrument cr1423a, and one additional work order was observed in march 2022 for the complaint failure mode reported.Complaints for results are within statistical control for the month of july 2022.The upper control limit was not breached, and trends were not identified associated with this defect.No new risks or hazards, or changes to existing risks/hazards, were identified as a result of this complaint.Per baltrm-maxinst-aph rev 22 the potential risk of the reported failure mode is false positive, which is an s3; reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd max¿ instrument, remanufactured there was discrepant results.No patient impact.The following information was provided by the initial reporter: customer reported discrepant results on 7/8, giardia was positive and on 7/10 repeated duplicates were negative.
|
| |
|