BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® NO ADDITIVE (Z) TUBES; SPECIMEN TRANSPORT AND STORAGE CONTAINER Back to Search Results
Model Number 366703
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported when using the bd vacutainer® no additive (z) tube the stopper pops out of the tube.The following information was provided by the initial reporter.The customer stated: the "non-bd caps are popping off.".
 
Additional Manufacturer Narrative
Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Additional Manufacturer Narrative
H6: investigation summary: bd had not received samples, but 1 photo was provided for investigation.The photo was reviewed and the indicated failure mode for stopper pop off could not be confirmed.The visual evaluation of the photo does show that the caps did come off the tubes; however, the caps that are seen in the photo are not the bd specified closure for this product.Per specification for catalog 366703, it is required to have a clear hemogard closure with a red stopper; the photo shows an orange closure.Therefore, the failure mode of stopper pop off cannot be confirmed.In addition, bd was unable to determine the specific lot number associated with this complaint; therefore, a review of the device history record could not be conducted.This complaint is unable to be confirmed.If additional information is made available, this complaint will be reopened to assess the level of investigation needed.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported when using the bd vacutainer® no additive (z) tube the stopper pops out of the tube.The following information was provided by the initial reporter.The customer stated: the "non-bd caps are popping off.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® NO ADDITIVE (Z) TUBES
Common Device Name
SPECIMEN TRANSPORT AND STORAGE CONTAINER
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15141382
Report Number1917413-2022-00446
Device Sequence Number5736373
Product Code FMH
UDI-Device Identifier50382903667036
UDI-Public50382903667036
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number366703
Device Catalogue Number366703
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2022
Supplement Date Received by Manufacturer08/19/2022
Initial Report FDA Received Date08/01/2022
Supplement Report FDA Received Date09/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-