• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL BOXED PATIENT CIRCUIT ASSY, 3100A, 850S; VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL BOXED PATIENT CIRCUIT ASSY, 3100A, 850S; VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number BOXED PATIENT CIRCUIT ASSY, 3100A, 850S
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluated by mfr: at this time, the suspect device has not been returned for evaluation.No root cause has been determined yet.The issue was with 3 circuits and two patients (it occurred repeatedly on one patient).Please see (b)(4).Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.".
 
Event or Problem Description
It was reported to vyaire medical that the secure connections on the wet side of the 29028-003 boxed patient circuit assy, 3100a, 850s disconnected while a patient was attached and being ventilated.This caused loss of pressure and stopped delivery.End user manually ventilated the patient until the circuit was changed.The customer confirmed that there was no patient harm associated with the reported event.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BOXED PATIENT CIRCUIT ASSY, 3100A, 850S
Common Device Name
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
510 technology dr
irvine IL 92618
Manufacturer (Section G)
PRODUCTOS UROLOGOS DE MEXICO S.A. DE C.V.
cerrada via de la produccion
no. 85, parque undustrial mex
mexicali,
MX  
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key15141840
Report Number8030673-2022-00262
Device Sequence Number5615061
Product Code LSZ
UDI-Device Identifier50190752159745
UDI-Public(01)50190752159745(10)0004155820(11)20200810
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P890057
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBOXED PATIENT CIRCUIT ASSY, 3100A, 850S
Device Catalogue Number29028-003
Device Lot Number0004155820
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/04/2022
Initial Report FDA Received Date08/01/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-