• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES SWITZERLAND SAR PROG VALVE INLINE W SG; CHPV WITH SG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES SWITZERLAND SAR PROG VALVE INLINE W SG; CHPV WITH SG Back to Search Results
Model Number 82-3832
Medical Device Problem Code Pressure Problem (3012)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 07/11/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A facility reported a hakim valve (id 823832) was implanted on (b)(6) 2022 with setting 100mmhg.After a month the patient did not recovered, then the physician set the pressure to 70mmhg however, hydrocephalus was not getting better.Then the physician reset the pressure to 50mmhg, the patient still had headache.The lumbar puncture for pressure testing showed the icp pressure was 300mmhg.The physician conducted a revision procedure and replaced with a new hakim valve on (b)(6) 2022 which was set at 120mmhg.After a week patient¿s symptom was relieved.
 
Additional Manufacturer Narrative
The hakim valve (id 823832) was not returned for evaluation after three good faith attempts (gfes) were made.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause of the reported issue could not be determined.However, a probable root cause for the reported complaint is due to biological debris and protein build up interfering with the valve mechanism.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event or Problem Description
N/a.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
The hakim valve (b)(4) was returned for evaluation.Failure analysis - the position of the cam when valve was received was on setting 50 mmh2o.The valve was visually inspected; no defect was noted.The valve was hydrated.The valve passed the test for programming, occlusion, reflux, siphon guard and pressure.No root cause could be determined as the technician could not confirm any problem with the valve at the time of investigation.The possible root cause for the issue reported by the customer, could be due to biological debris and protein build up interfering with the valve, at the time of investigation no function issues were noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROG VALVE INLINE W SG
Common Device Name
CHPV WITH SG
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key15142108
Report Number3013886523-2022-00343
Device Sequence Number5614572
Product Code JXG
UDI-Device Identifier10886704041825
UDI-Public(01)10886704041825(17)250228(10)4594292
Combination Product (Y/N)N
PMA/510(K) Number
K053107
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model Number82-3832
Device Catalogue Number823832
Device Lot Number4594292
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/11/2022
Supplement Date Received by Manufacturer11/08/2022
01/27/2023
Initial Report FDA Received Date08/01/2022
Supplement Report FDA Received Date11/18/2022
02/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/27/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-