|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that one opened box with ten packets of product code m8737 was returned with the packaging closed.Upon initial inspection, of the samples, no external damages were observed on the packets.In order to evaluate the conditions of the returned samples, the ten packets were opened, and no defects were detected.The swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related to damaged or breakage sutures and no defects were observed during the evaluation.A functional test was performed using an instron and the tensile strength force was above the minimum requirement.The event described could not be confirmed as the device performed without any defect noted.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.
|