It was reported that during an im nail procedure, doctor began drilling with the 2.0mm drill long w/ao qc without a guide and not in a very good position due to the type of approach being worked on.When going through the middle, the tip of the drill broke.Doctor was able to removed the tip from patient's bone.No pieces were left in the patient.Surgery was resumed, without a significant delay with a s+n back-up device.Patient was not harmed.
|
|
H3, h6.The device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.The patient's current condition is unknown, and the patient impact could not be determined based on the limited information provided.Should any additional clinical information be provided, this complaint will be re-evaluated.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This is a single use device, surgical technique or damage from misuse are likely potential factors that could contribute to the reported event.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|