• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CRE WIREGUIDED; DILATOR, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CRE WIREGUIDED; DILATOR, ESOPHAGEAL Back to Search Results
Model Number M00558410
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was used in the esophagus during a dilation procedure performed on (b)(6) 2022.During the procedure, the balloon was leaking water and a hole was visible on the balloon.The procedure was completed with another cre wireguided dilatation balloon.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Additional Manufacturer Narrative
Block h6: device code a041001 captures the reportable event of balloon pinhole in the esophagus.Block h10: investigation result: the returned cre wireguided dilatation balloon was analyzed and a visual examination that the balloon and catheter of the device found no damages.Functional analysis was performed and the balloon was inflated without a problem however, the balloon would not hold pressure due to a pinhole in the distal section of the balloon.Microscopic inspection found the balloon had a pinhole located approximately 16mm from the tip.With all the available information, boston scientific concludes the reported event of balloon pinhole was confirmed.It is possible that factors encountered during the procedure, such as the interaction with any surface during the procedure could create friction on the balloon and cause a pinhole.Therefore, the most probable root cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.A labeling review was performed and from the information available, this device was used per the instructions for use (ifu)/product label.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was used in the esophagus during a dilation procedure performed on (b)(6) 2022.During the procedure, the balloon was leaking water and a hole was visible on the balloon.The procedure was completed with another cre wireguided dilatation balloon.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE WIREGUIDED
Common Device Name
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15142745
Report Number3005099803-2022-04299
Device Sequence Number14835053
Product Code KNQ
UDI-Device Identifier08714729339342
UDI-Public08714729339342
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K110833
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model NumberM00558410
Device Catalogue Number5841
Device Lot Number0028125862
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/13/2022
Supplement Date Received by Manufacturer09/20/2022
Initial Report FDA Received Date08/01/2022
Supplement Report FDA Received Date10/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age62 YR
Patient SexMale
Patient Weight60 KG
-
-