BIOSENSE WEBSTER INC WEBSTER CS CATHETER WITH AUTO ID TECHNOLOGY; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
| Model Number |
D135304 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation findings code of ¿appropriate term/code not available¿ represents photo/video analysis.According to pictures provided by the customer, a foreign material was observed on the tip of the catheter, however the material looks like a human tissue and not an internal component, also, no damage was noted on the electrode.However, it is not possible to determine the origin of the foreign material based on the photo analysis.Although it is not possible to determine the foreign material, the customer complaint was confirmed due to the material attached to the tip.A manufacturing record evaluation was performed for the finished device 30732746m number, and no internal actions related to the reported complaint were identified.The customer complaint was confirmed from the photo analysis.However, the device has not been returned for evaluation.Since no device has been received, no product investigation can be performed.If the device is received in the future, the product analysis will be performed as appropriate in order to find the root cause of the complaint.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a cardiac ablation procedure with a webster cs catheter with auto id technology and a broken tip issue occurred.It was reported that the puller wire or some sort of internal component was protruding out of the tip of the catheter.The fellow had partially inserted it when the consultant asked for it to be pulled back so he could see the tip.There was no difficulty experienced while maneuvering the catheter, it was barely inserted.The damage did not result in any lifted or sharp rings.The catheter was swapped for a new catheter.There was no patient consequence, and the procedure was not extended greater than 30 minutes.
|
| |
|
Additional Manufacturer Narrative
|
|
On 01-sep-2022, the biosense webster, inc.Product analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).During a follow up inquiry, it was confirmed on 01-sep-2022, that the correct product lot number for this complaint is 30732845m.Therefore, the d4.Lot has been updated.
|
| |
|
Additional Manufacturer Narrative
|
|
The device evaluation was completed on 04-oct-2022.It was reported that a patient underwent a cardiac ablation procedure with a webster cs catheter with auto id technology and a broken tip issue occurred.It was reported that the puller wire or some sort of internal component was protruding out of the tip of the catheter.The fellow had partially inserted it when the consultant asked for it to be pulled back so he could see the tip.There was no difficulty experienced while maneuvering the catheter, it was barely inserted.The damage did not result in any lifted or sharp rings.The catheter was swapped for a new catheter.There was no patient consequence, and the procedure was not extended greater than 30 minutes.Device evaluation details: the product was returned to biosense webster inc.(bwi) for evaluation.Bwi conducted a visual inspection and deflection evaluation of the returned device.Visual analysis of the returned sample revealed a foreign material on tip area, the sample was sent to ft-ir (fourier transform infrared spectroscopy) for analysis, and no damage or anomalies were observed on the catheter.Then, the deflection testing was performed and observed that the tip deflects properly.The ft-ir results revealed that foreign material is primarily composed of a biological-based material, presumably human tissue.A manufacturing record evaluation (mre) was performed for the finished device 30732845m number, and no internal actions related to the reported complaint were identified.Based on the mre, the d4.Expiration date and h4.Device manufacture date have been updated.The event described could not be confirmed.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number:(b)(4).
|
| |
|