| Model Number |
ACU0T0 |
| Medical Device Problem Codes |
Crack (1135); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that during implantation of intraocular lens (iol), it can be pushed forward with considerable effort.Post implantation a hairline crack can be noted.The lens was not explanted and it was positioned to 12 o'clock.There was no patient impact.Additional information has been requested.There are three medical device reports associated with this event.This is 2 of 3.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in h.3., h.6.And h.10.The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.A qualified viscoelastic was indicated.The product investigation could not identify a root cause for the reported complaint.Follow up attempts were made for additional information.Follow up details indicated that the injector does do not exist anymore and the lens remains implanted.There is one other complaint in this lot.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|