Model Number ACU0T0
Medical Device Problem Codes Crack (1135); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that during implantation of intraocular lens (iol), it can be pushed forward with considerable effort.Post implantation a hairline crack can be noted.The lens was not explanted and it was positioned to 12 o'clock.There was no patient impact.Additional information has been requested.There are three medical device reports associated with this event.This is 2 of 3.
 
Additional Manufacturer Narrative
Additional information was provided in h.3., h.6.And h.10.The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.A qualified viscoelastic was indicated.The product investigation could not identify a root cause for the reported complaint.Follow up attempts were made for additional information.Follow up details indicated that the injector does do not exist anymore and the lens remains implanted.There is one other complaint in this lot.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM
Common Device Name
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15145136
Report Number1119421-2022-01686
Device Sequence Number17472900
Product Code KYB
UDI-Device Identifier00380652395328
UDI-Public00380652395328
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/13/2025
Device Model NumberACU0T0
Device Lot Number15372290
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/10/2022
Supplement Date Received by Manufacturer10/07/2022
Initial Report FDA Received Date08/02/2022
Supplement Report FDA Received Date10/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/15/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PROVISC OVD
Patient SexUnknown
-
-