• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON ET TUBE, SHER-I-BRONCH, LS, 35 FR; TUBE, TRACHEAL/BRONCHIAL, DIFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON ET TUBE, SHER-I-BRONCH, LS, 35 FR; TUBE, TRACHEAL/BRONCHIAL, DIFF Back to Search Results
Model Number IPN049000
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4) a device history record review was performed, and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event or Problem Description
It was reported that the doctor found connector boken during clinical setting before using on patient.No impact on patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON ET TUBE, SHER-I-BRONCH, LS, 35 FR
Common Device Name
TUBE, TRACHEAL/BRONCHIAL, DIFF
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key15145361
Report Number3003898360-2022-00328
Device Sequence Number20135584
Product Code CBI
UDI-Device Identifier14026704617446
UDI-Public14026704617446
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K180253
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date06/28/2026
Device Model NumberIPN049000
Device Catalogue Number5-16035
Device Lot Number73F2100793
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2022
Initial Report FDA Received Date08/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-