| Catalog Number |
814090100 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported a false negative reaction of a patient sample with ih-cell i-ii on ih-1000.The specificities of the missed antibodies were anti-k and anti-lea.Cell #2 of ih-cell i-ii was k and lea antigen positive and therefore supposed to react positively.The customer stated the antibodies were detected in the manual method.The anti-lea also showed a positive reaction in the repeat testing.The customer did neither provide the complaint sample ih- cell i-ii for investigational testing nor the patient sample that caused the false negative test result.But she announced to ship the patient sample for investigation.In the meantime, our quality control laboratory tested their retention sample of the supposedly defective lot.The testing was performed using ih-card ahg anti-igg lot #9203030, the same lot customer had used.Three different anti-k, three different anti-lea were used for the test as well as an anti-fya, anti-s and donor samples.All positive and negative reactions were correct.All antibodies were correctly detected.Additionally, the antigen k heterozygous positive cell of ih-cell i-ii (cell #2) was tested for potency in parallel with three other k antigen heterozygous positive reagent red blood cells (rrbc) and one antigen k homozygous rrbc.This testing was done manually.All k antigen heterozygous rrbs yielded the same titer.The testing confirmed that the k antigenicity of the claimed ih-cell i-ii lot was not weakened.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.Databases and log files of the affected instrument ih-1000 were collected.The analysis of the material for any issue relevant anomalies is still ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
This is our initial report on this incident.
|
| |
|
Event or Problem Description
|
|
The customer reported a false negative reaction of a patient sample with ih-panel plus 6 on their ih-1000 instrument.The specificity of the missed antibody was anti-k.Cell #14 of ih-panel plus 6 was k antigen positive and therefore supposed to yield a positive reaction.The customer stated the antibody was detected in the manual method.Furthermore, the customer stated, that the anti-k showed a positive reaction with the kell positive reagent red blood cells cell #1 and #8 of ih-panel 11.The customer did not provide the complaint sample ih- panel plus 6 for investigational testing but she provided a patient sample that had caused the false negative test result.Upon arrival on our premises the patient sample was completely hemolyzed.Even after centrifugation, no clear boundary between red blood cells and plasma was visible.While we were waiting for the patient samples to arrive, our quality control laboratory tested their retention sample of the supposedly defective lot.Three different anti-k were used for the test as well as an anti-fya, anti-s and donor samples.All positive and negative reactions were correct.All antibodies were correctly detected.Additionally, the antigen k heterozygous positive cell of ihpanel plus 6 (cell #14) was tested for potency in parallel with two other antigen k heterozygous reagent red blood cells (rrbc) and one antigen k homozygous rrbc.This testing was done manually.All k antigen heterozygous rrbs yielded the same titer.The testing confirmed that the k antigenicity of the claimed ih-panel plus 6 lot was not weakened.Due to the hemolysis of the patient sample a testing was not possible.The processed ih-cards forwarded by the customer could also not be analyzed, because the gel columns were completely dried out due to the lack of the sealing foil.Databases and log files of the affected ih-1000 were collected.The analysis of the received material for any issue relevant anomalies showed that the test was correctly performed.The distributed volume of serum and red blood cells was correct.The waiting time was a few seconds.In contrast to this, the waiting time of the ih-panel 11 where the anti-k was detected was 20 minutes.Based on the observed behavior, an igm component of the anti-k is highly suspected.Based on the investigation the complaint was classified as undetermined.The complaint sample was not available for investigation and the retention sample reacted as expected.Due to the grossly hemolysis the patient sample could not be tested.Comparative and potency tests demonstrated that the k antigenicity of the claimed ih-panel plus 6 lot was not weakened.Based on the observed behavior (positive reaction after a certain waiting time in the ih-1000), an igm component of the anti-k was highly suspected and would be a reasonable explanation for the negative reaction with ih-panel plus 6 at the customer's site.A general issue with the reagent red blood cells and ih-cards could be ruled out.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.Our initial report was issued for ih-cell i-ii but the customer later clarified that the incident occurred with ih-panel plus 6.
|
| |
|
Additional Manufacturer Narrative
|
|
This is our final report on this incident.
|
| |
|
Search Alerts/Recalls
|