• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. ROTATABLE CLIP FIXING DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. ROTATABLE CLIP FIXING DEVICE Back to Search Results
Model Number HX-110QR
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Full name of the facility is independent administrative institution national hospital organization chiba medical center.The number of times the device was used after delivery and the period of use are unknown.The brand and model number of the clip used is also not known.The device is returned and an evaluation completed for it.When the device was loaded with a test olympus clip there was no issue observed with loading and clipping.Upon inspection of the device, it was observed that the insertion part of the device was buckled at a position about 700 mm from the tip.The device hook was bent.There were no other abnormalities that could lead to the event which was reported.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
 
Event or Problem Description
As reported for this event by the customer, during a therapeutic hemostasis procedure, the clip was not release from the device, although the device slider was pulled and clicked.The device was able to clip but not release the clip.The device had to be forcibly removed and the procedure was completed with another similar device.There is no harm or adverse impact to the patient.The device was not inspected prior to use.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the bending of the hook was caused by the bending load applied when the connecting rod was removed after clipping or during reprocessing.The buckling of the coil sheath is likely to be caused by a bending load applied to the insertion site during reprocessing or insertion of the endoscope.However, the root cause could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTATABLE CLIP FIXING DEVICE
Common Device Name
ROTATABLE CLIP FIXING DEVICE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer Contact
masaharu hirose
2-248-1 okkonoki
kuroishi-shi, aomori 036-0-357
JA   036-0357
426422891
MDR Report Key15146326
Report Number9614641-2022-00083
Device Sequence Number6167593
Product Code PKL
UDI-Device Identifier04953170200205
UDI-Public04953170200205
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K013066
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHX-110QR
Device Lot Number74K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2022
Supplement Date Received by Manufacturer09/12/2022
Initial Report FDA Received Date08/02/2022
Supplement Report FDA Received Date10/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-