| Model Number |
HX-110QR |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Full name of the facility is independent administrative institution national hospital organization chiba medical center.The number of times the device was used after delivery and the period of use are unknown.The brand and model number of the clip used is also not known.The device is returned and an evaluation completed for it.When the device was loaded with a test olympus clip there was no issue observed with loading and clipping.Upon inspection of the device, it was observed that the insertion part of the device was buckled at a position about 700 mm from the tip.The device hook was bent.There were no other abnormalities that could lead to the event which was reported.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
|
| |
|
Event or Problem Description
|
|
As reported for this event by the customer, during a therapeutic hemostasis procedure, the clip was not release from the device, although the device slider was pulled and clicked.The device was able to clip but not release the clip.The device had to be forcibly removed and the procedure was completed with another similar device.There is no harm or adverse impact to the patient.The device was not inspected prior to use.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the bending of the hook was caused by the bending load applied when the connecting rod was removed after clipping or during reprocessing.The buckling of the coil sheath is likely to be caused by a bending load applied to the insertion site during reprocessing or insertion of the endoscope.However, the root cause could not be determined.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|