TELEFLEX MEDICAL PE ADULT-PED DRY/ WET LF 6/CS; BOTTLE, COLLECTION, VACUUM Back to Search Results
Model Number IPN028442
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been returned for investigation.
 
Event or Problem Description
During the decortication + pleurectomy procedure, thoracic drainage, pleur-evac, is placed and connected to the patient's tube, at which time it is identified that the hose is broken in the red connection , the connection closest to the pleur-evac.The device is changed and attention is continued with suction without problems.
 
Event or Problem Description
During the decortication + pleurectomy procedure, thoracic drainage, pleur-evac, is placed and connected to the patient's tube, at which time it is identified that the hose is broken in the red connection , the connection closest to the pleur-evac.The device is changed and attention is continued with suction without problems.
 
Additional Manufacturer Narrative
Qn#(b)(4).The device history record of batch number 74a2200234 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled & inspected according to our specifications.Manufacture date: 2022-01-11.Expiration date: 2025-01-10.No corrective action can be established at this moment since the product sample is not available for evaluation.Customer complaint cannot be confirmed based only on the information provided, to perform an investigation and determine the source of defect reported it is necessary to evaluate the sample involved on this complaint.
 
Additional Manufacturer Narrative
(b)(4).The device history record of batch number 74a2200234 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No nonconformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled & inspected according to our specifications.Manufacture date: 2022-01-11.Expiration date: 2025-01-10.One unit of catalog number a-6000-08lf (pe adult-ped dry/ wet lf 6/cs) lot 74a2200234 was received for analysis.Sample wasn't received in its original packaging.It was observed that one corner near to air leak meter is broken.The ats connector pn 152753 was inspected , and no broken issues were observed.After functional test the ats o-ring p/n 154589 was inspected , and no damage issues were observed.The ats connector and tubing were removed form unit, one side of tubing was applied air pressure (10 psi, the other side was clamped and ats connector was immerse into the water.No bubbles were observed that can confirmed a leak on connectors.Equipment tfm-0060.Nlc09684 due date 22-sep-23.Customer complaint cannot be confirmed based on the sample provided.The unit has a broken corner near to air leak meter.However, no leaks were observed , nor it can confirmed that such damage was originated during the manufacturing process.No issues were observed on ats connector p/n 152753 ("connection-red/blue") related with broken nor leak.
 
Event or Problem Description
During the decortication + pleurectomy procedure, thoracic drainage, pleur-evac, is placed and connected to the patient's tube, at which time it is identified that the hose is broken in the red connection , the connection closest to the pleur-evac.The device is changed and attention is continued with suction without problems.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PE ADULT-PED DRY/ WET LF 6/CS
Common Device Name
BOTTLE, COLLECTION, VACUUM
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key15146848
Report Number3011137372-2022-00146
Device Sequence Number17604671
Product Code KDQ
UDI-Device Identifier34026704631231
UDI-Public34026704631231
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K905768
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/10/2025
Device Model NumberIPN028442
Device Catalogue NumberA-6000-08LF
Device Lot Number74A2200234
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2022
Supplement Date Received by Manufacturer08/04/2022
09/05/2022
Initial Report FDA Received Date08/02/2022
Supplement Report FDA Received Date08/26/2022
09/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.; N/A.
Patient Age25 YR
Patient SexMale
-
-