• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 40" (102 CM) APPX 4.9 ML, ADMIN SET W/CHEMOLOCK¿ W/RED CAP, 20 DROP IN-LINE DRIP; CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 40" (102 CM) APPX 4.9 ML, ADMIN SET W/CHEMOLOCK¿ W/RED CAP, 20 DROP IN-LINE DRIP; CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM Back to Search Results
Model Number CL3020
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received for evaluation.However; the investigation is not yet complete.
 
Event or Problem Description
The event occurred on an unspecified date involving a 40" (102 cm) appx 4.9 ml, admin set w/chemolock¿ w/red cap, 20 drop in-line drip chamber, spiros®, bag hanger, drop-in red cap where the customer noticed a white sediment in the tubing.She added that when the tubing was cut open, the white matter can be removed which is a risk for patients.It was also reported that another product was opened (different lot number) and the tubing has the same issue.They started inspecting packages of cl3020 and multiple pieces (from a different lot number) had white sediment present.She also stated that the product package was sealed and not expired.A systemic treatment has a very high turnover of product as they have high usage.She would not think this package has sat for long but unable to give an exact time frame these items are stocked daily from their stores.There was no patient involvement and no patient harm.
 
Additional Manufacturer Narrative
Received one new list #cl3020, 40" (102 cm) appx 4.9 ml, admin set w/chemolock¿ w/red cap, 20 drop in-line drip chamber, spiros®, bag hanger, drop-in red cap; lot #5658373.The residual solvent in the inside of the tubing adjacent to the proximal bond port to the inline drip chamber is confirmed in the returned cl3020 administration set assembly.Subsequent investigation and testing showed that the residue of the dissolved white bond port solidified on the inside diameter of the clear tubing but was not present or observed when the fluid path was flushed and filtered after the tube was manipulated in the area of the residue.After cutting open the tubing and exerting heavy tube rubbing over the internal surface of the tube some of the white residue could flake off or be scraped off.The probable cause of the residual solvent is excessive solvent application during manual bonding in manufacturing.The lot history was reviewed, and no nonconformities were identified that may have contributed to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
40" (102 CM) APPX 4.9 ML, ADMIN SET W/CHEMOLOCK¿ W/RED CAP, 20 DROP IN-LINE DRIP
Common Device Name
CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
michael visocnik
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key15147182
Report Number9617594-2022-00222
Device Sequence Number12230945
Product Code ONB
UDI-Device Identifier00887709072923
UDI-Public(01)00887709072923(17)261101(10)5658373
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K131549
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCL3020
Device Catalogue NumberCL3020
Device Lot Number5658373
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2022
Supplement Date Received by Manufacturer08/08/2022
Initial Report FDA Received Date08/02/2022
Supplement Report FDA Received Date09/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-