Model Number TRF-385
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformance's noted.Device evaluation: visual analysis of the returned device identified: weight to the spec, crease fold, wear abrasion.Cloudy was observed after autoclave disinfection process.Based on the device analysis the final assessment is no issues found related with the manufacturing process.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv.
 
Event or Problem Description
Healthcare professional reported "capsular contracture baker grade iv" and anxiety against left device.The device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15147360
Report Number9617229-2021-55763
Device Sequence Number6167634
Product Code FTR
UDI-Device Identifier10888628008328
UDI-Public10888628008328
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/20/2021
Device Model NumberTRF-385
Device Catalogue NumberTRF-385
Device Lot Number2949119
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/23/2021
Initial Report FDA Received Date08/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/20/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-