• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 840 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 840 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 10049998
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic conducted an investigation based upon all information received.It was reported that during use on a patient, this 840 ventilator began to alarm and turned off.The device was available for evaluation.The service personnel (sp) inspected the ventilator and confirmed the reported issue of automatically turning off after turning on.The sp reported that the direct current (dc) source needed to be replaced.Additionally, the sp found an extended self-test (est) code and leak from the exhalation valve.The additional issue, required that the air proportional solenoid valve (psol) and exhalation valve be replaced.Multiple follow-up attempts, to the customer, have been made to obtain additional details with no response.The likely cause of the event was isolated in the field to a potential fault in dc source the manufacturing records for each device are thoroughly reviewed prior to release to ensure that it has met all medtronic quality specifications.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during use on a patient, this 840 ventilator began to alarm and turned off.It was reported that the patient died due to acute gastrointestinal bleeding and that the ventilator did not cause or contribute to the reported event of the patient's death.
 
Additional Manufacturer Narrative
Additional information added to section b5.Additional information added to initial report h3 device evaluation summary: multiple follow-up attempts, to the customer, have been made to obtain additional details with no response.Repair quotation has been submitted but not approved to date.The likely cause of the event was isolated in the field to a potential fault in dc power supply.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
Additional information as follows: it was reported that during use on a patient, this 840 ventilator began to alarm and turned off.The patient was found unresponsive.The patient was manually assisted with a resuscitator and then connected to another ventilator and death occurred after due to acute gastro intestinal bleeding, and it was determined that the ventilator did not cause or contribute to the patient's death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
EI 
Manufacturer (Section G)
COVIDIEN
micheal collins rd mervue
galway
EI  
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key15147556
Report Number8020893-2022-00132
Device Sequence Number5781947
Product Code CBK
UDI-Device Identifier10884521181045
UDI-Public10884521181045
Combination Product (Y/N)N
Initial Reporter CountryPE
PMA/510(K) Number
K970460
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10049998
Device Catalogue Number10049998
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/04/2022
Supplement Date Received by Manufacturer08/18/2022
Initial Report FDA Received Date08/02/2022
Supplement Report FDA Received Date09/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/08/2009
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexMale
-
-