• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD CRANIOTOMY KIT; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD CRANIOTOMY KIT; DISPOSABLE PERFORATORS Back to Search Results
Model Number 261230
Medical Device Problem Code Fail-Safe Did Not Operate (4046)
Health Effect - Clinical Code Brain Injury (2219)
Date of Event 07/06/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A facility reported a perforator did not disengage causing a tear in the dura matter.Suture was necessary to stop the bleeding of the superior sagittal sinus.The perforator was used in a surgery for meningioma.
 
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
The perforator was returned for evaluation: dhr - there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis - the perforator unit was inspected using the unaided eye.No anomalies were observed.Functional spring test and functional test were performed; no anomalies observed.Based off the failure analysis the complaint could not be confirmed.Root cause - he root cause is undetermined and was unable to be confirmed in the complaint evaluation.A review of the device history record and trending were performed as part of the evaluation.Product was received for analysis and the investigation could not confirm the complaint.Potential causes of failure include: user misuse.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRANIOTOMY KIT
Common Device Name
DISPOSABLE PERFORATORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key15148291
Report Number3014334038-2022-00174
Device Sequence Number5589254
Product Code HBF
UDI-Device Identifier10381780513629
UDI-Public10381780513629
Combination Product (Y/N)N
PMA/510(K) Number
K073336
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number261230
Device Catalogue Number261230
Device Lot Number6300030
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/20/2022
Supplement Date Received by Manufacturer09/29/2022
Initial Report FDA Received Date08/02/2022
Supplement Report FDA Received Date10/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-