| Model Number |
0684-00-0294-01 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Occupation: perfusionist.The product has been returned to the manufacturer, but is pending investigation.Once the investigation is completed a supplemental report with our findings will be submitted.Complaint record id #(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the customer could not insert the intra-aortic balloon (iab) over the guidewire.They noted that it would not thread.A new iab was inserted to continue therapy.There was no patient harm or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned with the membrane completely unfolded and traces of blood found on the exterior with the one-way valve attached.A guide wire was not returned for evaluation.The technician attempted to insert a laboratory 0.025¿ guide wire through the inner lumen and was able to successfully insert the guide wire.No obstructions were felt.The reported event cannot be confirmed by the evaluation.A lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Reference complaint #(b)(4).
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|