| Model Number |
1012276-20 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Separation (1562); Failure to Deflate (4060)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has been retained by the hospital and is not returning for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.The additional trek rx and the 4.0x12 xience skypoint referenced are being filed under separate medwatch report numbers.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat the mid right coronary artery (rca) and the distal rca.The patient had two unspecified xience skypoint stents implanted in (b)(6) of 2022.A 2.5x15mm trek balloon dilatation catheter (bdc) was used to pre-dilate between the two stents.Slow deflation of the bdc was noted, but had no further issues and was removed successfully.Optical coherence tomography (oct) confirmed there was a gap between the two previously implanted stents and a 4.0x12mm xience skypoint stent was deployed successfully with slow deflation of the device noted; however, the stent delivery system was removed successfully.A 4.0x38mm xience skypoint stent was attempted to be implanted proximally from the previously implanted xience skypoint in january; however, failed to be implanted as interaction with the previously implanted stent was felt.The 3.5x20mm trek balloon dilatation catheter was advanced, then inflated twice to 12 and 10 atmospheres.The physician attempted three times to pull negative, but the bdc would not deflate.It was decided to remove the bdc with the entire system as one unit.During removal the shaft separated and became lodged in the elbow area.The separated portion was attempted to be removed via snare but was difficult.Therefore, an emboshield nav6 was advanced and used to capture the very distal portion of the balloon just enough to remove the balloon which remained inflated.There was no adverse patient sequela but clinically significant delay was noted.No additional information was provided.
|
| |
|
Event or Problem Description
|
|
Subsequently, to the initial reports being filed it was noted that both 2.5x15 and the 3.5x20mm trek balloons had too much contrast; however, amount is unknown.It was confirmed by the account only one 4.0x28mm xience skypoint was implanted on (b)(6) 2022, and that the patient was experiencing angina, dyspnea and fatigue due to the thrombosis.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported failure to deflate; however, the reported difficulty removing the device, shaft separation, unexpected medical intervention, removal of foreign body and delay to treatment/therapy appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|