| Model Number |
VCP946H |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.(b)(4).The product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.Trade name - irgacare®.Active ingredient(s) ¿ triclosan.Dosage form ¿ suture/solid/parenteral.Strength ¿ 275 g/m.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a c-section procedure on an unknown date and suture was used.During the procedure, the needle popped off while suturing internal tissue.Another like device was used to complete the procedure.There were no adverse consequences for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint(b)(4).Date sent to the fda: 8/22/2022 this is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 investigational summary: one used needle of product code vcp946h was received to ethicon for evaluation.Visual inspection was conducted on the returned device.Upon visual inspection of the needle, the swage and attachment area were noted to be as expected, marks by the surgical instrument were noted on body needle and no suture remnant could be observed into the barrel hole.The suture was not received for evaluation to determine the assignable cause.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.No conclusion could be reached as to what caused the reported complaint.As part of quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.¿ trade name - irgacare® ¿ active ingredient(s) ¿ triclosan ¿ dosage form ¿ suture/solid/parenteral ¿ strength ¿ = 275 g/m.
|
| |
|
Search Alerts/Recalls
|
|
|